Ring, Sensor, and Special Installations
4-7
4. Position the ring so that the holes along the edge of the ring (the downstream
side) are facing you.
5. With the back of the ultrasonic mount facing you (the end with the small metal
peg pushed through the main plate), slide the mount onto the end of the ring’s
right side (the end without the welded metal band).
Adding the Sliding Ultrasonic Sensor Mount
6. Move the ultrasonic sensor mount to the left side of the ring.
7. Slide the end of the right side of the ring through the opening in the welded
band (on the left end of the ring) until there is about 10 cm or 4 in. of overlap.
8. Spread the overlapping sections of the ring until you are able to slide the ring
stabilizer and spreader mechanism screw into the gap, as shown in the
following illustration.
Ring Stabilizer Moved into Position