Connecting inputs and outputs – ADS Environmental Services Model 3600 530002 A2 User Manual
Page 49
4-16
ADS 3600 Flow Monitor O&M Manual
Connecting Inputs and Outputs
After the ring is prepared and the sensors have been attached, connect the cables to the
appropriate ports on the monitor lid. Ports on the lid are marked with symbols to
identify which sensor cables should be connected to which ports. Each sensor is
unique and will not fit into another sensor’s port.
The procedure for connecting cable connections to the lid ports follows.
l
Place the monitor in an upright position with the connectors facing upward.
l
Visually inspect each sensor connector and monitor port for damaged or
broken pins. If defective connectors are found in the lid port, replace the entire
monitor.
l
Visually inspect each sensor connector and monitor port for debris and
moisture. Clean off any debris and dry surface moisture.
l
Connect each sensor cable to the appropriate port. Tighten each connector in
a clockwise direction until it clicks and verify that it is seated correctly.
Doppler Velocity Sensor
Pressure Depth Sensor
Sampler
Communications Cable
Ultrasonic Level Sensor
Icons on the Monitor Lid Indicate Proper Cable Connections for Each Port
l
Seal any unused connectors with a cap.
l
Use waterproof splicing tape to seal around the sensor connectors and lid
ports. Apply Scotchkote
to seal.