Pressure depth subsystem, Troubleshooting 7-9 – ADS Environmental Services Model 3600 530002 A2 User Manual
Page 116
Troubleshooting
7-9
Problem
The velocity data readings are abnormally high.
Possible Cause
Wires on the P2 connector may be loose. The velocity sensor may be covered with silt.
The velocity sensor may be out of the flow. The flow may be reversed or slower than 0.15
m/s (0.5 ft/s).
Solution
Check the site conditions and relocate the velocity sensor if necessary.
If the problem is still unresolved, replace the monitor and send the old one to ADS for
service.
Pressure Depth Subsystem
Use the following tips to help resolve problems with the pressure depth subsystem.
Warning:
Prior to replacing a sensor, contact an ADS trained technician for
further diagnosis.
Problem
The temperature reading from the pressure depth sensor shows a value of -273
°
C
(-459
°
F).
Possible Cause
The cable connecting the pressure mouse to the monitor may be broken. The pressure
sensor may be faulty. The pressure depth board is not responding or may be faulty.
Solution
Replace the pressure sensor with a known good sensor and send the faulty one to ADS for
repair.
If the problem is still unresolved, replace the monitor and send the old one to ADS for
service.
Problem
The pressure depth subsystem data readings are consistently incorrect.
Possible Cause
The coefficients are incorrect.
Solution
Enter the correct coefficients and reactivate the monitor.
Problem
The pressure depth subsystem data readings are consistently off by up to 7.62 cm (3 in.),
but temperature readings are accurate.
Possible Cause
The pressure sensor has an offset that drifts over time.
Solution
Check the condition of the pressure dryer tube.
Adjust the depth data by applying a delta offset to the readings until they match the values
obtained during confirmation.
Problem
The pressure sensor temperature readings are incorrect, but depth data is accurate.
Possible Cause
The temperature sensor may be faulty.
Solution
Contact an ADS trained technician for further information about replacing the pressure
sensor.
If the problem is still unresolved, replace the monitor and send the old one to ADS for
service.