beautypg.com

4 cleaning the instrument and accessories, Cleaning the instrument and accessories, Section 4.6.4 – Metrohm 873 Biodiesel Rancimat Manual User Manual

Page 88

background image

4.6 Determinations

■■■■■■■■■■■■■■■■■■■■■■

80

■■■■■■■■

873 Biodiesel Rancimat

The determination is automatically saved in the database
Repos.mrd. The channel is ready for starting a new determination.

2 Clean vessels and accessories

Note

Always clean the vessels and accessories directly after each deter-
mination
(see Chapter 4.6.4, page 80).

4.6.4

Cleaning the instrument and accessories

1 Clean the measuring vessels and accessories

Clean used measuring vessels after pouring off the measuring
solution with acetone or 2-propanol. Pre-clean with washing-up
liquid in the case of severe contamination.

Rinse the product thoroughly with distilled water.

Clean the measuring vessel covers (5-

16

), the PTFE tubes

(5-

17

) and the electrodes (5-

18

) with acetone or 2-propanol and

rinse with distilled water. Pre-clean with washing-up liquid in the
case of severe contamination.
Remove the protective ring (5-

19

) to be able to clean the electrodes.

2 Clean the reaction vessels and accessories

Dispose of used reaction vessels and air tubes and use new
reaction vessels and air tubes for the next measurement.

Clean the reaction vessel covers with acetone or 2-propanol and
rinse with distilled water. Pre-clean with washing-up liquid in the
case of severe contamination.

Then heat the reaction vessel covers in the drying oven for 2
hours at 80 °C.

Note

Replace the reaction vessel covers if they are not tightly on the reac-
tion vessel anymore or the material is obdurate.

3 Clean the tubings

Clean the Iso-Versinic

®

tubings (5-

2

) with acetone or 2-propanol

and rinse with distilled water. Pre-clean with washing-up liquid in
the case of severe contamination.