beautypg.com

Metrohm 873 Biodiesel Rancimat Manual User Manual

Page 20

background image

3.2 Mounting accessories

■■■■■■■■■■■■■■■■■■■■■■

12

■■■■■■■■

873 Biodiesel Rancimat

2 Mount drying flask

Caution

Do not fill the hot molecular sieve directly into the drying flask after
regeneration, as otherwise the plastic filter on the filter tube will
melt.

Wait until the molecular sieve has cooled down before filling.

Fill the molecular sieve into the drying flask (4-

2

).

Screw the filter tube (4-

4

) on the lower side of the drying flask cover

(4-

1

) into the opening marked (above) with a dot.

Screw the drying flask cover with mounted filter tube onto the dry-
ing flask and insert in the flask holder (4-

3

) on the rear of the

Biodiesel Rancimat.

Screw the one end of the FEP tubing 250 mm (4-

6

) onto that open-

ing on the drying flask cover at the bottom of which the filter tube
is located.

Screw the other end of the FEP tubing onto the From flask con-
nection (4-

7

) on the rear of the Biodiesel Rancimat.

Screw the one end of the second FEP tubing 250 mm (4-

5

) onto

the second opening on the drying flask cover.

Screw the other end of the second FEP tubing onto the To flask
connection (4-

9

).

Note

The molecular sieve serves for adsorbing interfering oxidizing gases
as well as water from the sucked in air. You can regenerate it in the
drying oven at +140…+180 °C for 24 to 48 h (see Chapter 5.3, page
156)
.

3 Mount the FEP tubing for the supply air

Screw the one end of the FEP tubing 130 mm (4-

11

) onto the Air

out connection (4-

10

).

Screw the other end of the FEP tubing onto the Air/N

2

in connec-

tion (4-

12

).