beautypg.com

Metrohm 873 Biodiesel Rancimat Manual User Manual

Page 73

background image

■■■■■■■■■■■■■■■■■■■■■■

4 Operation

873 Biodiesel Rancimat

■■■■■■■■

65

Evaluate induction time

on | off (Default value: on)

Determination of the induction time (see "Induction time", page 66).

Evaluate stability time

on | off (Default value: off)

Determination of the stability time (see "Stability time", page 67).

Note

If you select this evaluation mode, you always need to calibrate the con-
ductivity measuring cell beforehand.

Delta Kappa

Conductivity change for the determination of the stability time.

Range

1 ... 400 µS/cm

Options for induction time

Evaluation delay

Waiting time until starting the curve evaluation.