7 Technical specifications

873 Biodiesel Rancimat

■■■■■■■■

167

7.5

GLP test set

External tempera-
ture sensor

Type

Pt100 casing element

⌀1.9 mm, in 4-conductor technology; DIN IEC 751

Class B

Calibration
accuracy

0…90 °C: ±0.02 °C
90…220 °C: ±0.03 °C

Max. deviation
of the measured
temperature
from the true
value (50…220
°C)

±0.3 °C

Test resistor

10 k

Ω

7.6

RS-232 interface

Plug

Sub-D plug, 9-pin (male)

Basic settings

9600 Baud, 8 Bit, 1 stop bit, no parity, XON/XOFF

7.7

Mains connection

Voltage

2.873.0014: 220…240 V ± 10 %
2.873.0015: 100…120 V ± 10 %

Frequency

50…60 Hz

Power consump-
tion

450 W

Fuse

Diameter 5 mm, length 20 mm
100…120 V: 2.0 ATH (slow-acting)
220…240 V: 4.0 ATH (slow-acting)