3 Confirm or reject results

Test passed: The deviation of the measured gas flow values
is within the specified tolerance.

Test failed: The deviation of the measured gas flow values
is greater than the specified tolerance.

Confirm the results with [Finish].

or

Reject the results with [Cancel].

4.8.5

GLP results

Printing the GLP certificate

Note

To be able to print a test certificate, you have to switch on the GLP mon-
itoring (see "Activating/Deactivating GLP monitoring", page 138).

For the selected instrument, you can print a GLP certificate, which contains
the following information:

Print date, serial number of the instrument, results of the last GLP tests (with
date, test person and test results).