background image

■■■■■■■■■■■■■■■■■■■■■■

4 Operation

873 Biodiesel Rancimat

■■■■■■■■

69

4.5.2.4

"Formulas" tab

You can enter values for further automatic calculations with the induction
and stability time on this tab. The settings here do not have any effect on
the measuring sequence or the evaluation. After finishing the measurement,
a post-calculation or recalculation is possible at any time in the determination
overview (see Chapter 4.7.5, page 117).

Formula 1 for induction time:

on | off (Default value: off)

If this option is switched on, the automatically determined induction time
will be calculated with the entered values. The result will be saved under the
name entered in the field "R1".

Formula 2 for stability time:

on | off (Default value: off)

If this option is switched on, the automatically determined stability time will
be calculated with the entered values. The result will be saved under the
name entered in the field "R2".