4.8 GLP functions

■■■■■■■■■■■■■■■■■■■■■■

148

■■■■■■■■

873 Biodiesel Rancimat

Click on the menu item Tools

GLP test Conductivity... and

confirm with [Next >].
The following dialog window appears:

Select the option Use standard solution.

Under Conductivity of standard solution, enter the conductivity
in µS/cm and confirm with [Next >].

c(KCl) = 1 mmol/L can be used as a standard solution. This is derived
from the optionally available conductivity standard 6.2301.060
(KCl 0.1 mol/L) through dilution with distilled water. The conductivity
of this diluted solution is:

Temperature Conductivity

18 °C

19 °C

20 °C

21 °C

22 °C

23 °C

24 °C

25 °C

127 µS/cm

130 µS/cm

133 µS/cm

136 µS/cm

138 µS/cm

141 µS/cm

144 µS/cm

147 µS/cm

The conductivity is measured sequentially at all channels.

After the last measurement is completed, the dialog window with the
test results appears.