4 Operation
873 Biodiesel Rancimat
■■■■■■■■
63
per block
The 4 channels of a block can be started and stopped jointly.
Stop criteria
Time
on | off (Default value: off)
Time until the automatic stop of the determination.
Range
0.1 ... 9999 h
Conductivity
on | off (Default value: off)
Conductivity limit value. If this value is reached, the determination is ended
automatically.
Range
1 ... 400 µS/cm
Endpoint(s)
on | off (Default value: off)
Automatic stop of the determination, as soon as every active channel has
reached all endpoints (induction time and stability time).
Table 2
Overview of stop criteria
Time
Conductivity
Endpoint(s)
Analysis stops...
not automatically.
✔
after the time entered, independently of reach-
ing the induction/stability time (endpoint).
✔
after reaching the conductivity entered, inde-
pendently of reaching the induction/stability
time (endpoint).
✔
after reaching the induction and stability time
(endpoint) defined on the Evaluation tab. If the
induction and stability time are evaluated, both
endpoints have to be found.
✔
✔
after reaching the first criterion (time or conduc-
tivity), independently of reaching the induction/
stability time (endpoint).
✔
✔
after reaching the first criterion (time or end-
point). If the time entered has been reached,
there is no further measuring until the induction/
stability time (endpoint) has been reached.