162

■■■■■■■■

873 Biodiesel Rancimat

Problem

Cause

Remedy

The conductivity measuring
cell is contaminated.

Check measuring cell (see Chapter
4.8.4.2, page 146)
and, if necessary,
clean.

See also: The induction times
are not reproducible for multi-
ple determinations.

The induction time is
longer/shorter than
expected.

The temperature is not selec-
ted correctly.

Ensure that the correct method for the
determination has been selected.

Check that Temperature and
Delta T have been specified correctly
in the method (see Chapter 4.5.2.1,
page 60)
.

See also: The induction times
are not reproducible for multi-
ple determinations.

The stability time is lon-
ger/shorter than expec-
ted.

Delta Kappa is not defined cor-
rectly.

Ensure that the value of Delta Kappa
defined in the method is correct (see
Chapter 4.5.2.2, page 64)
.

See also: The stability times are
not reproducible for multiple
determinations.

See also: The induction time is
longer/shorter than expected.

The measurement
curves are extremely
noisy.

The air supply for the measur-
ing solution is directed to the
conductivity measuring cell.

Loosen tubing adapter (5-

11

) on the

measuring vessel cover, turn the PTFE
tube (5-

17

) in such a way that the air is

no longer directed to the measuring cell
and fix in this position.

Gas bubbles adhere to the
conductivity measuring cell
during measurement.

Ensure that the measuring cell is clean
and free of fat. If necessary, clean
thoroughly (see Chapter 4.6.4, page
80)
.

In some cases ultra pure water con-
tains a large proportion of dissolved
air which escapes during measure-
ment. In this case, degas the ultra pure
water before the measurement for 5
to 10 min in a vacuum.