4 Operation

873 Biodiesel Rancimat

■■■■■■■■

79

4 Insert and connect reaction vessels

Note

The temperature defined in the method must be reached, before you
insert the reaction vessels, i.e. the frame of the button [Heat to xx
°C]
must be green.

Seal the channels not used with the sealing plugs or empty reaction
vessels for protection against impurities.

Insert the prepared reaction vessels into the openings of the heating
block.

Connect the Iso-Versinic

®

tubings, which are fastened to the reac-

tion vessel cover, to the tubing adapter M8 / olive (5-

11

) of the

measuring vessel covers.

Screw the FEP tubings 250 mm (5-

1

) to the thread adapters (5-

3

)

of the reaction vessel covers and the air supply connections (3-

3

) of

the Biodiesel Rancimat.

4.6.3

Starting the determination

Before you can carry out the determination, the channels you intend to use
must be ready(see Chapter 4.6.7, page 82).

1 Start determination

Start of an individual determination: Click on the button
[Start] of the corresponding channel.

Start of all channels of Block A: Click on the button .

Start of all channels of Block B: Click on the button .

After the determination has been started, the live curve is displayed
for every channel:

As soon as a stop criterion defined in the method (see table 2, page
63)
has been reached, the determination is ended: