Metrohm 873 Biodiesel Rancimat Manual User Manual
Page 122
4.7 Results
■■■■■■■■■■■■■■■■■■■■■■
114
■■■■■■■■
873 Biodiesel Rancimat
Extrapolation of several results
Proceed as follows to carry out an extrapolation of several results:
1 Open dialog window
■
Open the Results window and the database (see "Displaying the
determination overview", page 83).
■
Select the required determinations (see "Selecting determina-
tions", page 86).
■
Click on the menu item Determination
▶ Extrapolation.
The dialog window Extrapolation with the extrapolation values which
the program has calculated from the selected results (determinations)
appears.
2 Select result and target temperature
■
Click on the [Settings] button.
The following dialog window appears:
■
Select the result which is to be used for extrapolation (Induction
time, Stability time, etc.).
■
Enter the Target temperature.
■
Confirm the entries with [OK].
Note
The time determined for a particular sample by extrapolation to low
target temperatures (e.g. 25 °C) is a purely theoretical value. The
interrelation between this time and e.g. the perishability of the sam-
ple must be determined experimentally.
- 915 KF Ti-Touch (382 pages)
- 800 Dosino (53 pages)
- 767 Calibrated Reference (23 pages)
- 940 Professional IC Vario ONE/SeS/Prep 2 (54 pages)
- 754 Dialysis Unit (49 pages)
- 815 Robotic Soliprep for LC (76 pages)
- Vision Manual (207 pages)
- tiamo 2.1 Manual (1532 pages)
- 825 Lab Link (37 pages)
- 808 Titrando (70 pages)
- 902 Titrando (52 pages)
- 756 KF Coulometer (163 pages)
- 756 KF Coulometer (162 pages)
- 940 Professional IC Vario ONE/LPG (98 pages)
- 850 Professional IC Anion MCS Prep 3 (154 pages)
- 850 Professional IC Anion MCS Prep 3 (152 pages)
- 904 Titrando (58 pages)
- 850 Professional IC Anion MSM-HC MCS Prep 2 (150 pages)
- 930 Compact IC Flex Oven/ChS/Deg (47 pages)
- 872 Extension Module Liquid handling (64 pages)
- 814 USB Sample Processor (90 pages)
- 814 USB Sample Processor (91 pages)
- 940 Professional IC Vario (43 pages)
- Vision – Tutorial (40 pages)
- 799 GPT Titrino (242 pages)
- 889 IC Sample Center (68 pages)
- 761 Compact IC (228 pages)
- 851 Titrando (100 pages)
- 748 DH Sample Changer (32 pages)
- 940 Professional IC Vario ONE/SeS/HPG (51 pages)
- 896 Professional Detector – Amperometry (62 pages)
- 877 Titrino plus (139 pages)
- 881 Compact IC pro – Anion (129 pages)
- 940 Professional IC Vario ONE/ChS/HPG (112 pages)
- 930 Compact IC Flex Deg (41 pages)
- 840 PC Control 5.0 / Touch Control (351 pages)
- 940 Professional IC Vario ONE/Prep 1 (45 pages)
- 776 Dosimat (42 pages)
- 717 Sample Changer (36 pages)
- 815 Robotic USB Sample Processor XL (113 pages)
- 815 Robotic USB Sample Processor XL (114 pages)
- 940 Professional IC Vario ONE/SeS/PP (126 pages)
- 838 Advanced Sample Processor Installation Instructions (109 pages)
- 700 Dosino (55 pages)
- 719 S Titrino (152 pages)