beautypg.com

Metrohm 873 Biodiesel Rancimat Manual User Manual

Page 62

background image

4.4 Calibration functions

■■■■■■■■■■■■■■■■■■■■■■

54

■■■■■■■■

873 Biodiesel Rancimat

Note

You can find the calibration data (T1 to T3) for Metrohm sensors on
the certificate enclosed with the sensor or on the label fastened on
the sensor cable.

The following dialog window appears:

2 Wait for measurement

Wait until the message Measured temperature is now stable!
appears in the message box.

The following three values are displayed below the message box:

Block temperature

Temperature which has been measured in the
heating block with the internal temperature sen-
sor.

Temperature of ext. sensor

Temperature which has been measured in the
reaction vessel with the external temperature
sensor.

Delta T

Temperature correction which is determined
from the difference between the block temper-
ature and the temperature of the ext. sensor.