4.8 GLP functions

■■■■■■■■■■■■■■■■■■■■■■

144

■■■■■■■■

873 Biodiesel Rancimat

The following dialog window appears:

4 Insert GLP test set and start measurement

Insert the prepared test set (see "Preparing the GLP test set", page
141)
in the channel of Block A to be checked.

Connect the temperature sensor to the Pt100 connector on the rear
of the 873 Biodiesel Rancimat.

In the dialog window GLP test for unit with serial no. x, click
on [Next >].

The test measurement is started.

As soon as the temperature defined in the method has been reached,
the following message appears:

Ext. temperature

The temperature measured with the exter-
nal temperature sensor.

Measuring time

The time elapsed since the start of the tem-
perature measurement.