2 replacing fuses, 3 mains cable and mains connection, Replacing fuses – Metrohm 873 Biodiesel Rancimat Manual User Manual

Page 26: Mains cable and mains connection, See chapter 3.3.2

background image

3.3 Mains connection

■■■■■■■■■■■■■■■■■■■■■■

18

■■■■■■■■

873 Biodiesel Rancimat

3.3.2

Replacing fuses

Two fuses 4 A (slow-acting) for 115 V or 2 A (slow-acting) for 230 V are
installed in the fuse holder (3-

6

) of the 873 Biodiesel Rancimat as standard.

Warning

Ensure that the instrument is never put into operation with fuses of
another type, otherwise there is a risk of fire!

Proceed as follows to replace defective fuses:

1 Pull out the mains cable

Pull the mains cable out of the mains connection socket of the
Biodiesel Rancimat.

2 Remove the fuse holder

Use a screwdriver to loosen and completely pull out the fuse holder
(3-

6

) above the mains connection socket.

3 Replace fuses

Carefully take the defective fuses out of the fuse holder and replace
them with two new fuses suitable for the set mains voltage of the
type TH (slow-acting, with high switching capacity):

– 115 V 4 A (TH) Metrohm No.: U.600.0022
– 230 V 2 A (TH) Metrohm No. U.600.0107

4 Insert the fuse holder

Push the fuse holder back into the instrument until it latches into
place.

3.3.3

Mains cable and mains connection

Mains cable
The mains cable optionally supplied for the instrument

6.2122.020 with plug SEV 12 (Switzerland)

6.2122.040 with plug CEE(7), VII (Germany, …)

6.2122.070 with plug NEMA 5-15 (USA, …)

is three-core and provided with a plug with grounding pin. If another plug
has to be mounted, the yellow/green conductor (IEC standard) must be
connected to the protective ground (protection class I).