3 Installation
873 Biodiesel Rancimat
■■■■■■■■
17
Proceed as follows to mount the collection tube:
1 Mount the exhaust collection tube
■
Insert the exhaust collection tube with both nozzles into the collec-
tion tube holders (3-
) on the Biodiesel Rancimat in such a way that
the connection to the exhaust air removal is located to the rear.
2 Connect the measuring vessels
■
Screw the tubing adapter M8 / olive (5-
) into the opening Out
) of the measuring vessel cover.
■
Connect the one end of the Iso-Versinic
®
) to the tubing
adapter M8 / olive.
■
Insert the other end of the Iso-Versinic
®
tubing into the correspond-
ing opening on the collection tube.
■
Seal the unused openings on the collection tube with the enclosed
plugs E.400.0010.
3 Connect the exhaust collection tube
■
Connect a suitable tubing to the connection of the exhaust collec-
tion tube and connect this to an active suction device (e.g. water
jet pump).
3.3
Mains connection
Warning
There is a risk of fire if the instrument is operated with an incorrect mains
fuse!
Follow the regulations below for the mains connection.
3.3.1
Checking the mains voltage
Before switching the 873 Biodiesel Rancimat on for the first time, check
whether the mains voltage indicated on the type plate (3-
) corresponds to
the mains voltage present. If this is not the case, please contact Metrohm
Service.