3 Installation

873 Biodiesel Rancimat

■■■■■■■■

15

15 Connector plug

16 Measuring vessel cover (6.0913.130)

Contains integrated conductivity measuring
cell.

17 PTFE tube (6.1819.080)

For supplying the air to the measuring solu-
tion.

18 Electrode

19 Protection ring

20 Glass measuring vessel (6.1428.030)

Proceed as follows to mount the measuring and reaction vessel:

1 Mount the measuring vessel cover

Insert the PTFE tube (5-

17

) from above into the opening In (5-

13

)

of the measuring vessel cover.

Screw the tubing adapter M8 / olive (5-

11

) into the opening In of

the measuring vessel cover.

Place the measuring vessel cover (5-

16

) on the measuring vessel

(5-

20

).

2 Mount the reaction vessel cover

Insert the air tube (5-

5

) from below into the connection (5-

7

) of the

reaction vessel cover.

Place the O-ring (5-

6

) over the upper end of the air tube.

Screw the thread adapter M8 / M6 (5-

3

) gently into the connection

(5-

7

) and, at the same time, press the air tube against the

thread adapter M8 / M6.

Now fix the air tube onto the reaction vessel cover by firmly pulling
the thread adapter M8 / M6.

If determinations are carried out with highly foaming samples,
clamp the foam barrier (5-

9

) onto the air tube.

Place the reaction vessel cover on the reaction vessel.

Warning

The foam barrier can melt if it projects too deeply into the heating block.

Ensure that the foam barrier (5-

9

) is at least 7 cm above the base of the

reaction vessel (5-

10

).