4.7 Results
■■■■■■■■■■■■■■■■■■■■■■
108
■■■■■■■■
873 Biodiesel Rancimat
Directory
Directory in which the report has been saved.
Description
The following warning messages are possible for incorrectly executed deter-
minations:
Temperature not reached: setpoint temperature at start of the determi-
nation not reached.
Missing measuring points (start time): measuring points are missing
from the given time (in s) onwards.
Missing measuring points (end time): measuring points are present
again from the given time (in s) onwards.
Manual Stop: the determination was stopped manually with [Stop].
Stop after communication error: Automatic stop of the determination
because of interrupted connection between the Biodiesel Rancimat and the
PC.
Value
Value for warning (e.g. temperature, time).
4.7.3
Graph and reevaluation
You can either open a graph of an individual determination (single
graph) or have graphs of several determinations displayed above one
another (multiple graph). In addition, you can open the so-called live
graph of a determination, in the dialog window of which you can manually
perform reevaluations of measuring curves.
You have the following options for reevaluation of graphs:
■
Setting the induction time manually
■
Setting the stability time manually
■
Setting tangents manually