background image

4.5 Methods

■■■■■■■■■■■■■■■■■■■■■■

62

■■■■■■■■

873 Biodiesel Rancimat

Table 1

Approximate values for determining Delta T

Setpoint temperature

Delta T

80 °C

90 °C

100 °C

110 °C

120 °C

130 °C

140 °C

150 °C

160 °C

+0.7 °C

+0.8 °C

+0.8 °C

+0.9 °C

+1.0 °C

+1.1 °C

+1.1 °C

+1.2 °C

+1.2 °C

Gas flow

Gas flow through the sample. The gas flow can only be set for both blocks
jointly. The value of the method loaded for block A is always decisive.

Range

7 ... 25 L/h

Use internal pump

on | off (Default value: on)

Switch on/off gas supply with internal pump. If the internal pump is switched
off, an external gas supply must be connected to the Air/N

2

in connection

(see Chapter 3.2.2, page 13).

Start options

Start delay

Waiting time until starting the measurement.

Range

0 ... 1000 min

Start mode

Mode for starting and stopping the channels.

Selection

per channel | per 2 channels | per block

per channel
All channels can be started and stopped individually.
per 2 channels
2 channels each can be started and stopped jointly. The [Start] button
in the control window changes automatically for this.