4 Operation

873 Biodiesel Rancimat

■■■■■■■■

145

5 Carry out measurement for further channels

Warning

The aluminum cylinder of the GLP test set is very hot. Do not touch!

As soon as the message appears with the prompt to insert the
external temperature sensor together with the measuring insert in
the next selected channel, rapidly carry out the change.

Carry out measurement for each of the selected channels.

After the last measurement is completed, the dialog window with the
test results appears.

6 Confirm or reject results

Test passed: The deviation of the measured temperatures
is within the specified tolerance.

Test failed: The deviation of the measured temperatures is
greater than the specified tolerance.

Confirm the results with [Finish].

or

Reject the results with [Cancel].