6 Troubleshooting

873 Biodiesel Rancimat

■■■■■■■■

159

6 Troubleshooting

6.1

Problems

Problem

Cause

Remedy

The pump is louder
than normal.

The air flow is blocked some-
where before or after the
pump.

Check dust filter (3-

20

) and, if neces-

sary, replace.

Check filter tube (4-

4

) on the drying

flask cover for blockages and, if nec-
essary, free with a gentle tap.

Examine FEP tubings (2-

6

), (3-

13

),

(3-

14

) and (3-

17

) for blockages and,

if necessary, replace.

Extra air is sucked in at some-
where else other than the dust
filter. There is a leak some-
where in the system before the
pump.

Check FEP tubings (3-

13

), (3-

14

) and

(3-

17

) for cracks, kinks etc. and

tighten. If necessary, replace.

Place the drying flask cover (3-

10

)

correctly on the drying flask and screw
tight.

No air flow can be
detected in the reaction
vessel (it does not bub-
ble), although the
pump is running.

The air supply is blocked.

Remove FEP tubing (5-

1

) from the

thread adapter (5-

3

). A slight air flow

must be discerned here. If this is not
the case, please contact Metrohm
Service.

Check FEP tubing (2-

6

) for blockages.

If necessary, clean or replace.

Check thread adapter (5-

3

) and air

tube (5-

5

) on the reaction vessel cover

for blockages. If necessary, clean or
replace.

The FEP tubing for the air sup-
ply is defective.

Check FEP tubing (2-

6

) for cracks, kinks

etc. If necessary, replace.

The FEP tubing for the air sup-
ply is not connected correctly.

Tighten FEP tubing (2-

6

) on both sides.

The air tube (5-

5

) does not

immerse in the sample.

Push reaction vessel cover (5-

8

) down

as far as the limit stop.

Use more sample.