Metrohm 873 Biodiesel Rancimat Manual User Manual
Page 126

4.7 Results
■■■■■■■■■■■■■■■■■■■■■■
118
■■■■■■■■
873 Biodiesel Rancimat
The dialog window Recalculation appears.
2 Adjust the parameters
■
Activate the options Recalculate induction time and Calcula-
tion active.
■
Under Evaluation delay and Evaluation suppression start/
end, enter the required values.
■
Under Evaluation sensitivity, enter the required value.
The software suggests a value with which the highest maximum of
the second derivative is accepted as the induction time. If there is
no suggestion then the highest maximum has already been accep-
ted.
The default value "1" will lead to the correct result for most appli-
cations. Smaller values will cause the acceptance of smaller maxi-
mums of the second derivative. Higher values require a higher
maximum.
■
Confirm the entries with [OK].
Note
A detailed description of the parameters can be found at "Recalcu-
late induction time", page 120.
Recalculating the stability time
Proceed as follows to recalculate the stability time:
1 Open dialog window
■
Select the determination which is to be recalculated (see "Selecting
determinations", page 86).
■
Click on the menu item Determination
▶ Recalculation.
The dialog window Recalculation appears.
2 Adjust the parameters
■
Activate the options Recalculate stability time and Calculation
active.
■
Under Delta Kappa enter the required value and confirm with
[OK].