beautypg.com

Metrohm 873 Biodiesel Rancimat Manual User Manual

Page 126

background image

4.7 Results

■■■■■■■■■■■■■■■■■■■■■■

118

■■■■■■■■

873 Biodiesel Rancimat

The dialog window Recalculation appears.

2 Adjust the parameters

Activate the options Recalculate induction time and Calcula-
tion active
.

Under Evaluation delay and Evaluation suppression start/
end
, enter the required values.

Under Evaluation sensitivity, enter the required value.
The software suggests a value with which the highest maximum of
the second derivative is accepted as the induction time. If there is
no suggestion then the highest maximum has already been accep-
ted.
The default value "1" will lead to the correct result for most appli-
cations. Smaller values will cause the acceptance of smaller maxi-
mums of the second derivative. Higher values require a higher
maximum.

Confirm the entries with [OK].

Note

A detailed description of the parameters can be found at "Recalcu-
late induction time", page 120
.

Recalculating the stability time

Proceed as follows to recalculate the stability time:

1 Open dialog window

Select the determination which is to be recalculated (see "Selecting
determinations", page 86)
.

Click on the menu item Determination

Recalculation.

The dialog window Recalculation appears.

2 Adjust the parameters

Activate the options Recalculate stability time and Calculation
active
.

Under Delta Kappa enter the required value and confirm with
[OK].