background image

■■■■■■■■■■■■■■■■■■■■■■

4 Operation

873 Biodiesel Rancimat

■■■■■■■■

71

Factor

Parameter for the calculation of the standard time, which indicates the factor
by which the time changes per 10 °C. This value can be determined through
the extrapolation (see Chapter 4.7.4, page 113) of results, which have been
recorded at different temperatures.

Range

0.001 ... 100

Target temperature

Target temperature for the calculation of the standard time.

Range

20 ... 300 °C

4.5.2.6

"Documentation" tab

This tab contains parameters for the automatic report output.

Include in report...

Header

on | off (Default value: on)

Output of the header with company settings (see ""Company settings"
tab", page 133)
, date of the printout, sample identifications, instrument
information, determination date, user, cell constant and serial number.