3 transmitter output connections (section #1), Transmitter output connections (section #1) -13, Warning – CiDRA SONARtrac PW VF-100 User Manual
Page 79

20832-01 Rev 05
Page 6-13
6.3.4.3
Transmitter Output Connections (Section #1)
User supplied data output cable size AWG 22 to AWG 16 (0.326 mm
2
to 1.31 mm
2
) is installed through a cable gland in the furthest left hole
on the transmitter housing and attached to the appropriate terminal
block connection points. The hole is sized for a 3/4 inch NPT or M25
cable gland fitting (1-1/16 inch (25.4 mm) hole).
WARNING
For Class I, Division 2 applications, Power Entry and
Inputs/Outputs are not certified as Non-incendive and must be
installed in accordance with National Electrical Code
requirements.
WARNING
For ATEX Class I, Zone 2 applications, Power Entry and
Inputs/Outputs are not certified as Non-incendive and must be
installed in accordance with EN60079-14. See SONAR PROCESS
MONITORING SYSTEM SUPPLEMENT FOR ATEX ZONE 2
SAFETY for additional info and requirements