CiDRA SONARtrac PW VF-100 User Manual
Page 183

20835-01 Rev 04
Page 9-9
Following the entering of ‘Pipe Size’ press the ‘BACK’ key and
‘→Pipe Size’ will be displayed on Line 2. At this point the user can
either press the
‘ENTER’ key to re-enter the ‘Pipe Size’ menu or
press the
↓ key to move to the next menu item.
Pressing the
↓ key will next display ‘→Fluid Properties’ on Line 2.
Press the ‘ENTER’ key and ‘→Specific Gravity’ will appear on
Line 3 and the current value on Line 4. To change the Line 4 value
press ‘ENTER’ and use the arrow keys to enter the new value.
Water at 25 ºC (0.997) is the default. Appendix E lists values for
water at various temperatures. Once the new value has been
entered on Line 4, press
‘ENTER’.
If no change is made (or after a change in Specific Gravity has
been made) pressing the
↓ key will next display ‘→Viscosity (Pa
s)
’ on Line 3 and the current value on Line 4. To change the Line 4
value press ‘ENTER’ and use the arrow keys to enter the new
value. Water at 25 ºC (8.9008 e
-04
) is the default. Appendix E lists
values for water at various temperatures. Once the new value has
been entered on Line 4, press ‘ENTER’.
Pressing the
↓ key will next display ‘→Calibration’ on Line 2.
Press the
‘ENTER’ key and ‘→C0’ will appear on Line 3 and a
numerical value on Line 4. Press the ‘ENTER’ key and use the
arrow keys to enter the calibration factors that will accompany the
sensor. Once the ‘C0’ value is entered press the ‘ENTER’ key to
store that value to memory. Line 3 will show ‘→C0’; press the ↓ key
to move to
‘→C1’, press ‘ENTER’, and use the arrow keys to enter
‘C1’ values. Once the values are entered, press ‘ENTER’ and
press the
↓ key to move to ‘→C2’ and enter those values and press
‘ENTER’ and the ‘BACK’ key.
The display will show ‘→Calibration’ on Line 2. Press the ↓ key to
scro
ll to ‘→Flow Direction’ on Line 2. If it is necessary to change
the flow direction (if the sensor was installed with the Flow Direction
arrow on the band opposite the actual flow direction within the pipe
or the process flow has changed direction) press t
he ‘ENTER’ key
and toggle the
↑ or ↓ to change flow direction. Once changed press
the
‘ENTER’ key.
Press the
↓ key to scroll to ‘→Set Date/Time’ on Line 2. Press the
‘ENTER’ key and the current saved date and time will be displayed
on Line 4 of the display. Use the arrow keys (
↑ and ↓ to scroll
through the values and
← and → to change character) to enter the
date and time. Note: Time is in 24 hour format. Once the date and
time have been set press the
‘ENTER’ key.
Press the
↓ key to scroll to ‘→Set Date Format’ on Line 2. Press
the ‘ENTER’ key and the current date format will be displayed on
Line 4 of the display. Use the arrow keys (
↑ and ↓) to scroll to the