8.5.5
Flow Cutoff Range (Systems with Flow)
The flow cutoff range ‘Low End’ and ‘High End’ are used to set the
low end and high end, respectively, of flow rates that will be displayed
on the transmitter. Flow under or over the set points will be displayed
as ‘
8.5.6
Totalizer (Systems with Flow)
Selections under this menu are used to configure the totalizer
functions.
8.5.6.1
Units
Units submenu is used to select the transmitter displayed units.
Options are ‘gal’ (gallons), ‘m
3
’
(cubic meters), ‘VF Vol Units’
(volumetric flow display units), ‘l’ (liters) ‘ft
3
’
(cubic feet).
8.5.6.2
Lowcut Enable
Lowcut Enable turns the lowcut function on or off.
8.5.6.3
Lowcut
Lowcut sets the flow values that will no longer be used for totalization.
Flow below this value will not be totalized.
8.5.6.4
Multiplier
Multiplier selects the multiplier applied to the displayed totalized units.
Selections of ‘M’ (x 1,000,000), ‘k’ (x 1,000), ‘1’ (x 1) are available.
Note:
This selection does not affect the pulse output when ‘Totalizer’
is selected. Set the pulse output multiplier in the ‘Pulse’ setup menu.
8.5.6.5
Reset
Reset sets the totalizer display to zero.
8.5.6.6
Input (Systems with Flow and GVF)
Input is used to select ‘volumetric flow’ or ‘true liquid flow’ to
totalize.
8.5.7
Wr Protect Mode
When this is ‘Enabled’ no other parameters can be changed. The
user mus
t ‘Disable’ this option before making any menu changes.
Default is ‘Disable’.