Figure 44, Sensor cover bolt tightening sequence -39, Caution – CiDRA SONARtrac PW VF-100 User Manual
Page 57

20831-01 Rev 05
Page 5-39
10
6
2
4
8
11
7
3
1
5
9
12
Alignment bolts
Figure 44 Sensor Cover Bolt Tightening Sequence
Note: The gaskets on the cover assembly will compress and conform
to the pipe surface during installation. Upon removal the gasket will
relax a little and will still provide proper sealing if re-installed at the
same location from where it was removed.
CAUTION
The sensor cover gasket on fiberglass covers should be
replaced, if the assembly is moved to another pipe, to minimize
the potential for leakage into the cover. Refer to rework
instructions (RI-0001) for information.
Following installation of the cover fasteners remove the alignment
bolts.