System faults (page 5) – CiDRA SONARtrac PW VF-100 User Manual
Page 265

20838-01 Rev 09
Page J-6
System Faults (page 5):
Item
No
PROBLEM
POSSIBLE
CAUSE
DIAGNOSTIC
METHOD
DIAGNOSTIC
RESULT
COMMENTS
9
The Display
is Frozen
(reads same
value all the
time)
Transmitter left in
Idle Mode
Is Heart Beat Indicator
Turning in Lower Right
Corner
Yes
– Heart Beat
Indicator is Turning
Acquire Snapshot, Upload to secure Website,
and contact Customer Support.
No- Heart Beat Indicator
is not turning
Cycle power to transmitter
Transmitter is
hung up
Cycle power to
transmitter. Does the
display update?
Yes
– Display updates
Meter is properly functioning
No
– Display is still
frozen
Acquire Snapshot, Upload to secure Website,
and contact Customer Support.
10
The Display
Reads
Preamp
Failure and
there is no
flow reading
Possible faulty
Preamp or bad
connection
Verify wiring and
connectors are
properly connected
No
– wiring / connector
fault found
Correct fault
Yes
– all wiring and
connectors is correct
Acquire Snapshot, Upload to secure Website,
and contact Customer Support.