Table 11 – CiDRA SONARtrac PW VF-100 User Manual
Page 127

20833-01 Rev 04
Page 7-33
Table 11
GVF Transmitter Menu Tree Software Release 04.02.XX
Level 1
Level 2
Level 3
Range
Description
Basic Config
Sensor
Serial #
0000000
Serial number of sensor
band
Pipe Size
ID/Wall
ID: 0.1- 100 in (2.54-
2540mm)
Wall 0
– 100 in (0 to
2540 mm)
Pipe inner diameter and
wall thickness
Size/Sched
2 to 36" size; schedule
Pipe size & schedule
OD/Wall
OD: 0.1- 300 in (2.54-
7260mm)
Wall 0
– 100 in (0 to
2540 mm)
Pipe outer diameter and
wall thickness
Pipe Material
SST, CS, PVC, Custom
Enters the pipe
modulus, kilo-Pascal
(kPa)
Fluid
Properties
Spec Gravity
0 - 999999
Enters the specific
gravity; default water at
25 °C and 14.7 psia
SOS (ft/s)
0 - 999999
Sound speed in media
of interest; ft/sec;
default water at 25 °C
and 14.7 psia
Pressure
+/- 0-999999
Process pressure; PSIg,
Barg, kPag
Temperature
-999 to +999C
-1766 to 1830F
Process temperature;
°C or F
Pressure Sel
Fixed, Sensor #1 or #2,
Protocol
Allows for selecting a
fixed pressure input or
using the inputs of
sensor #1 or #2 or using
Modbus input
Temperature
Sel
Fixed, Sensor #1 or #2
Allows for selecting a
fixed temperature input
or using the inputs of
sensor #1 or #2 or using
Modbus input
Altitude
-50,000 to +50,000
Process pipe altitude
above / below sea level;
feet or meters
Set
Date/Time
MM/DD/YY
HH:MM:SS
Based on Format
Current date/time;
adjust for daylight
savings as required
Set Date
Format
US(MM/DD/YY)
EURO (DD/MM/YY)
ISO8601(YY-MM-DD)
Format of date display
on transmitter