Figure 17, Orientation (clocking) of sensor cover -21 – CiDRA SONARtrac PW VF-100 User Manual
Page 39

20831-01 Rev 05
Page 5-21
It is helpful to have a second person available to help when installing
the cover assembly. When the sensor head is installed on a
horizontal pipe, the sensor cover should be installed such that the
transmitter cable connector socket is located within the 105° arcs
shown in the following figure. Do not install the cover with the
transmitter cable connector socket installed downward. (An electrical
pre-amplifier board is mounted on the inside of the cover.) The ability
to route the sensor band to sensor cover cable may dictate cover
orientation. The cable gland on the sensor head to transmitter cable
connector, when installed, will face away from the cover.
Figure 17 Orientation (Clocking) of Sensor Cover
If the sensor head is installed on a vertical pipe, the cable connector
should be located so it is facing downward.
Upper cover
assembly
Lower cover assembly
Transmitter cable
connector socket