Table 10 – CiDRA SONARtrac PW VF-100 User Manual
Page 119

20833-01 Rev 04
Page 7-25
Table 10
VF Transmitter Menu Tree Software Release 04.02.XX
Level 1
Level 2
Level 3
Range
Description
Basic Config
Sensor
Serial #
0000000
Serial number of sensor
band
Pipe Size
ID/Wall
ID: 0.1- 100 in (2.54-
2540mm)
Wall 0
– 100 in (0 to
2540 mm)
Pipe inner diameter and
wall thickness
Size/Sched
2 to 36" size; schedule
Pipe size & schedule
OD/Wall
OD: 0.1- 300 in (2.54-
7260mm)
Wall 0
– 100 in (0 to
2540 mm)
Pipe outer diameter and
wall thickness
Fluid
Properties
Liq Spec
Gravity
0 - 999999
Enters the liquid specific
gravity; default water
0.997
Viscosity
0.0000 e-38 to
9.9999 e+38
Enters the liquid
viscosity in Pa-sec;
default water 8.9008e
-4
Calibration
Coefficients
C0 term
0.0000 e-38 to
9.9999 e+38
First term coefficient
C1 term
0.0000 e-38 to
9.9999 e+38
Second term coefficient
C2 term
0.0000 e-38 to
9.9999 e+38
Third term coefficient
Flow
Direction
Forward, Reverse
Sets direction of flow
relative to ‘Flow
Direction’ arrow on
sensor band
Set
Date/Time
MM/DD/YY
HH:MM:SS
Based on Format
Current date/time; Note:
Unit does not adjust for
daylight savings.
Manually adjust as
required.
Set Date
Format
US(MM/DD/YY)
EURO (DD/MM/YY)
ISO8601(YY-MM-DD)
Format of date display
on transmitter