Evaluation procedure

Eppendorf BioSpectrometer

®

kinetic

English (EN)

12.6.3

Conversion to amounts of dye

The quantity (dimensions) of dye in the entire sample volume is calculated in the

process results/More

calculations method step.

M = calculated total amount (mass) of dye in the sample tube. Unit: pmol.

C = dye concentration calculated from the measurement. Unit: pmol/μL.

V

S, total

= total volume of the sample in the sample tube; entered by the user under

More calculations. 

Unit: μL.

12.7

Dual wavelength

For methods of the

Dual Wavelength group absorbances that were measured at two wavelengths can be

calculated with each other before the calculated absorbance is evaluated further with the factor or
standard.

To determine the calculated absorbance a division or subtraction evaluation can be defined in the
parameters: