Application parameters

Methods

Preprogrammed and freely programmable methods for all measuring
and evaluation procedures:

• Absorbance measurements with one or more wavelengths, scans
• Nucleic acids and proteins, OD600, dye methods (parallel

measurement of biomolecule and dye marking)

• Methods with evaluation via factor, standard and standard series
• Dual wavelength procedure with subtraction and division

evaluation

• Kinetic method: End point, two-point, linear regression

Method-dependent evaluation

Absorbance, concentration via factor and standard.
Concentration via standard series:

• Linear regression
• Nonlinear regression (2nd and 3rd degree polynoms)
• Spline evaluation
• Linear interpolation (point-to-point evaluation)

Absorbance calculations via subtraction and division
Additional data for nucleic acids: Ratios 260/280 and 260/230; molar
concentration, total yield
Additional data for dye methods: FOI (frequency of incorporation,
labeling density)
Scans: zoom, peak evaluation
Kinetics: Subsequent modification of the time window for the
regression evaluation

Method memory

>100 method programs

Measured value memory and
calibration memory

Memory for >1 000 results with all data of the results evaluation and
standard evaluation, sample number, sample name, date and used
parameter set of the method program.
(The number of saved results depends on the number of saved
methods.)