Eppendorf BioSpectrometer kinetic User Manual

Page 36

background image

Methods
Eppendorf BioSpectrometer

®

kinetic

English (EN)

36

Correct A260 1

Selection:
on | off

Only for the

Dye labels method group.

Correction of the influence of the dye spectrum on the
absorbance with the measuring wavelength of the biomolecule
(260 nm or 280 nm). Some of the dye spectra have a low
absorbance at 260 and 280 nm. These absorbances distort the
calculations for the nucleic acids or the proteins of these
methods. To minimize the distortion correction factors are
used, as long as these are known for the corresponding dyes.
When the parameter is switched on, the correction factor is
imported from the

General Method Parameters/Dyes function.

Correct A 280 1

Selection:
on | off

Only for the

Dye labels method group.

For an explanation see the description of the

Correct A260 1

parameter above.

Dye 2 active

Selection:
on | off

Only for the

Dye labels method group.

Option to measure a second dye at the same time. Application:
Labeling a biomolecule with two dyes.

Dye 2

Selection:
List of dyes that are
stored in the

General

Method Parameters/
Dyes function.

Only for the

Dye labels method group when measuring 2 dyes.

Selection of the second dye (see

Dye 1 parameter).

Correct A260 2

Selection:
on | off

Only for the

Dye labels method group when measuring 2 dyes.

Analog to parameter

Correct A260 1.

Correct A280 2

Selection:
on | off

Only for the

Dye labels method group when measuring 2 dyes.

Analog to parameter

Correct A280 1.

Show scan

Selection:
on | off

Display of a scan (absorbance wavelength graph) in addition to
the result of the sample measurement.

Start λ

Value input:
Wavelength in nm.
Range: 200 to 830 nm.

Starting wavelength for recording the scan.

Stop λ

Value input:
Wavelength in nm.
Range: 200 to 830 nm.
Value must be higher
than the value for 
Start λ.

Stopping wavelength for recording the scan.

A260/A280

Selection:
on | off

Only for nucleic acids.
Display of the A260/A280 ratio in addition to the result of the
sample measurement.

A260/A230

Selection:
on | off

Only for nucleic acids.
Display of the A260/A230 ratio in addition to the result of the
sample measurement.

Parameter

Entry