6.4.5

Process results

The sample measurement is followed by two optional steps in the method sequence:

process results and

print & export.

In the

process results step, you can postprocess the results for some methods. Example: Changing the

spectra section of a scan.

As for the result display, you can navigate between the sample results of the measurement series with the

and cursor keys and select results for postprocessing.

Tab. 6-3:

Options: Overview

Option

Explanation

Available in methods

Zoom

Change the axis limit of the absorbance
wavelength graph to limit the view to
enlarged sections of the graph.

Generally, all methods for which the
Scan parameter is offered and has
been activated.

Multi λ

Scan

Nucleic acids

Proteins direct UV

Dye labels

Bacterial density

Dual wavelength

More calculations

Convert concentration results into molar
concentrations and (after entering the
volume) into total amounts.

Nucleic acids

Dye labels (with nucleic acids as
biomolecule)

Peak detection

Identifying peaks in absorbance wavelength
spectra.

Scan

Linear regression

Modifying time windows for the evaluation of
a kinetic via linear regression.

In principle, all kinetics methods for
which the "Linear regression"
measuring procedure was used.

Single λ - continuous

Simple kinetics

Advanced kinetics

Options for postprocessing are presented on the two
softkeys at left. In this example: [Zoom] and [More
Calculations].