• For efficient temperature control, the volume of the measuring solution in the cuvette

should not project beyond the edge of the cuvette holder.

• To speed up the measuring procedure during series measurements, you can pre-cool

cuvettes with reagents in a thermostat outside the BioSpectrometer before inserting the
cuvette in the cuvette holder and adding the sample.

• When switching from a method with temperature control to one without temperature

control, please note that the temperature of the cuvette holder will slowly drift back to
room temperature. The results of the method without temperature control could be
affected by this.