Operation

Eppendorf BioSpectrometer

®

kinetic

English (EN)

Measured values typical for Eppendorf for temperature control in closed cuvettes, with closed cuvette shaft
covers, are shown in the following tables. The temperature was measured in the measuring solution; the
ambient temperature was 24.5°C.

Tab. 5-1:

The temperature to be set for temperature control of a measuring solution

Tab. 5-2:

Duration for temperature control of a measuring solution

Cuvette type/measuring volume

Target temperature

Temperature to be set in the
method parameters

Quartz glass macrocuvette, 
1 500 μL

25°C

25°C

30°C

30.4°C

37°C

37.7°C

Plastic macrocuvette, 
1 000 μL

25°C

25°C

30°C

30.4°C

37°C

37.7°C

Quartz glass semimicrocuvette,
500 μL

25°C

25°C

30°C

30.4°C

37°C

37.7°C

Quartz glass ultra-micro cuvette,
60 μL

25°C

25°C

30°C

30.3 °C

37°C

37.6°C

Cuvette type/measuring volume

Initial temperature and target
temperature

Temperature control duration

Quartz glass macrocuvette, 
1 500 μL

25°C - 37°C

Approx. 7 min

37°C - 25°C

Approx. 11 min

25°C - 30°C

Approx. 7 min

Plastic macrocuvette, 
1 000 μL

25°C - 37°C

Approx. 13 min

37°C - 25°C

Approx. 19 min

Quartz glass semimicrocuvette,
500 μL

25°C - 37°C

Approx. 7 min

37°C - 25°C

Approx. 12 min

Quartz glass ultra-micro cuvette,
60 μL

25°C - 37°C

Approx. 7 min

37°C - 25°C

Approx. 9 min

25°C - 30°C

Approx. 5 min