Evaluation procedure

Eppendorf BioSpectrometer

®

kinetic

English (EN)

The factor is either entered directly as a parameter during the "Factor" evaluation procedure or calculated
during the "Standard" evaluation procedure (evaluation with a standard concentration):

F = calculated factor.

C

S

= concentration of the standard (enter as parameter).

A

S

= measured absorbance of the standard.

If multiple measurement (2 or 3 replicates) has been programmed for the standard, the average value is
calculated from the measured absorbance values and inserted as A

S

.

12.3

Evaluation with standard curve/line

If evaluations are made with more than one standard, the following evaluation procedures for the standard
curve/line can be selected with the [Curve fit] in the

measure standards/new method step:

For the regression procedure, one can select that the regression line (regression curve) goes through the
zero point.

Evaluation procedure

Description

Minimum required number of
standard points

Linear interpolation

Linear point-to-point connection in the
absorbance concentration graph of the
standard evaluation.

2 standards minimum.

Linear regression

Polynome regression for first degree
polynomial.

3 standards minimum.

Quadratical regression

Polynome regression for second degree
polynomial.

4 standards minimum.

Cubical regression

Polynome regression for third degree
polynomial.

5 standards minimum.

Spline interpolation

Interpolation via natural cubic splines.

3 standards minimum.

S

S

A

C

F