2 photometry method description, 1 absorbance method group, Photometry method description 6.2.1 – Eppendorf BioSpectrometer kinetic User Manual
Page 30: Absorbance method group
Methods
Eppendorf BioSpectrometer
®
kinetic
English (EN)
30
You can create new methods in all folders using
In Favorites, you can create your own folders (e.g., to allocate folders to specific people), and rename and
delete the folders.
Tab. 6-2:
Softkeys in method selection
Copied or cut methods can be added to a different folder under Favorites, or added to the original folder
under a new name. Use the cursor keys to navigate to the
Methods column of the desired folder and press
[paste] for adding the method.
6.2
Photometry method description
The preprogrammed methods and method templates are described in this section.
6.2.1
Absorbance method group
Single λ
• Absorbance measurement on a wavelength.
• No subsequent evaluation.
Single λ - continuous
• Repeated absorbance measurement on a wavelength.
• Parameters can be entered for temperature control between 20 and 42°C (presetting: 37°C).
• Entry of parameters for total time and interval time for the measuring points. A premature stop cannot
be made during the measurement.
• Evaluation as kinetic via
linear regression. A subsequent change cannot be made to the time frame for
the evaluation.
• The measured values are displayed in an absorbance time graph.
• To subsequently evaluate measured values as linear regression kinetics, press the [Next >] softkey and
go to the
process results method step.
Multi λ
• Absorbance measurement at two to six wavelengths.
• No subsequent evaluation.
Scan
• Absorbance wavelength spectra measurement via a defined wavelength range.
• Display of wavelength and absorbance in the spectrum by navigation with a wavelength cursor.
• The spectra section can be modified using 3 different zoom variants.
• Peak detection possible.
[Cut] and [Paste]
Cut and paste methods.
[Copy] and [Paste]
Copy and paste methods.
[Delete]
Delete methods.
[Rename]
Rename methods.