Methods
Eppendorf BioSpectrometer

®

kinetic

English (EN)

48

Bacterial
density

Results display:

• Calculated result with absorbance at

the measuring wavelength.

• If activated in the parameters: Scan.

Navigate between the measuring

points in the graph with and .

Basic

main group

Factor,
standard

Analog to Protein direct UV (see above)

Results display:

• Concentration result with absorbance

at the measuring wavelength.

Calibration
curve

Analog to Proteins (with reagent) (see above)

Results display:

• Concentration result with absorbance

at the measuring wavelength.

• Graph of the standard evaluation with

plotted sample result.

Simple
kinetics

Prior to measurement:

Check parameters method step:
Display temperature control. When the
temperature set in the parameters has
been reached and the display changes
from

Tempering to Ready, you can go

to the measurement.

Results display:

• Concentration result with absorbance

or A/min with the measuring
wavelength.

• Only for the "linear regression"

measuring procedure: graph with
absorbance time display and plotted
regression line.
Navigate between the measuring

points in the graph with and .

Additional data ([Data] softkey):
Dependent on the measuring procedure:

• Absorbance time value pairs for the

first and last measuring points
("endpoint" method: only one
measuring point).

• Quality parameters for linear

regression (omitted for kinetics with a
slope below 0.050 E/min).

Method group Results display

Explanation