Maintenance

Eppendorf BioSpectrometer

®

kinetic

English (EN)

8.3.1.1 Checking photometric accuracy

1. Select the

Spectrometer unit function in the

Device calibration group and confirm with enter.

2. Select whether you want to check the wavelength

systematic error or photometric accuracy, and
confirm with

enter. Press [Next >] to switch to

the measurement.

3. Follow the instructions on the device display and

start by measuring the A0 blank filter, and then
the first test filter A1.
The device measures the test filter 15 times at
9 wavelengths, and then shows the average value
and the VC values for the series of measurements
for all 9 wavelengths.