4 photometric properties, 5 incubation, Photometric properties – Eppendorf BioSpectrometer kinetic User Manual
Page 86: Incubation, 4 photometric properties 11.5 incubation
Technical data
Eppendorf BioSpectrometer
®
kinetic
English (EN)
86
11.4
Photometric properties
11.5
Incubation
Measuring principle
Single beam absorption spectrophotometer with reference beam
Light source
Xenon flash lamp
Monochromator
Holographic aberration-corrected concave grating
Beam receiver
CMOS photodiode array
Wavelengths
200 nm to 830 nm
Wavelength selection
Method-dependent, freely selectable
Spectral bandwidth
≤ 4 nm
Smallest step size
1 nm
Systematic wavelength error
±1 nm
Random wavelength error
≤ 0.5 nm
Photometric measuring range
0 A to 3.0 A at 260 nm
Reading accuracy
ΔA = 0.001
Random photometric error
≤ 0.002 at A = 0
≤ 0.005 (0.5 %) at A = 1
Systematic photometric error
±1 % at A = 1
Stray light component
< 0.05 %
Adjustable temperature range
20°C to 42°C
Smallest step size
0.1°C
Maximum temperature
uncertainty (in relation to the
temperature in the sample)
Systematic temperature error
±0.2 °C at 25 °C to 37 °C
Random temperature error
±0.15 °C at 25 °C to 37 °C