4 measure samples: results displays, Measure samples: results displays – Eppendorf BioSpectrometer kinetic User Manual

Page 45

background image

45

Methods

Eppendorf BioSpectrometer

®

kinetic

English (EN)

6.4.4

Measure samples: Results displays

This section contains a display of typical results displays for all method groups and an overview of
additional results data, which can be accessed using the [Data] softkey.

Method group Results display

Explanation

Absorbance

main group

Single λ

Results display:

• Absorbance at the measuring

wavelength

• Only for dilution or with cuvettes other

than 10 mm: additional display of the
absorbance value before the
conversion.

Single λ -
continuous

Prior to measurement:

Check parameters method step:
Display temperature control. When the
temperature set in the parameters has
been reached and the display changes
from

Tempering to Ready, you can go

to the measurement.

During the measurement:

• You can prematurely cancel the

measurement using the [Stop] softkey.

Results display:

• Graph with absorbance time display

and plotted regression line.

• Value for "A/min" calculated from the

linear regression.

Additional data ([Data] softkey):

• Absorbance time value pairs for the

first and last measuring points.

• Quality parameters for linear

regression

• Navigate between the measuring

points in the graph with and .

• If necessary, change the time frame

for evaluation with linear regression in
the

process results method step.