beautypg.com

Eppendorf BioSpectrometer kinetic User Manual

Page 82

background image

Troubleshooting
Eppendorf BioSpectrometer

®

kinetic

English (EN)

82

During the kinetics
measurement, the
temperature was
outside of the
allowable range.

• Ambient temperature outside the

specified range.

• The temperature control is faulty.

 Measure the sample at ambient

temperature within the specified
range (15°C to 35°C). If the
warning nevertheless appears,
contact Eppendorf Service.

Absorbance at the
measuring
wavelength is too
high.

• Turbidity of the measuring solution.
• Optical surfaces of the cuvette are soiled.
• Cuvette has been inserted into the

cuvette shaft facing the wrong direction.

• Too high absorbance of measuring

solution.

 Measure again considering the

possible causes.

The calculated
result is negative.

• Measuring solution not prepared

correctly.

• The incorrect factor has been entered

(wrong algebraic sign).

 Measure again considering the

possible causes.

At least one of the
results is negative.

• For methods with several results (e.g.,

Dye labels).

• Measuring solution not prepared

correctly.

• The incorrect factor has been entered

(wrong algebraic sign).

 Measure again considering the

possible causes.

The result has more
than 6 pre-decimal
places.

• Very high sample concentration.
• Concentration unit does not match the

expected range of the sample
concentrations.

 Dilute sample and measure again.
 Change the concentration unit

(Parameter

Unit) and measure

again.

The result is more
than 5 % outside of
the standard
concentration
range.

• For methods with evaluation via standard

curves (nonlinear evaluation method):
The sample result is more than 5 %
outside of the standard concentration
range.

 Remeasure the sample under

conditions under which the result
is within the range of the standard
concentrations (dilute sample,
modify standard concentrations
and remeasure).

• Calculation not

possible
because of
division by zero.
Absorbance
result is zero.

• Calculation

error. Division
by zero.

• The evaluation required dividing by an

absorbance result with the value of
"zero". This is not mathematically
permissible.
Examples: Calculation of a factor at
one-point calibration; calculation of a
260/280 ratio with nucleic acid
measurements.

 Check the reagents and samples

used and repeat the measurement.

Calculation not
possible because of
division by zero.
Absorbance result
or parameter
formula

b is zero.

• An absorbance result was divided by a

"zero" value during the evaluation of a
Division type method (Dual wavelength
method group). This is not
mathematically permissible.

 Check the reagents and samples

used and repeat the measurement.

 Do not enter "zero" as a value for

the Formula

b parameter.

Problem

Cause

Solution