Eppendorf BioSpectrometer kinetic User Manual
Page 82
Troubleshooting
Eppendorf BioSpectrometer
®
kinetic
English (EN)
82
During the kinetics
measurement, the
temperature was
outside of the
allowable range.
• Ambient temperature outside the
specified range.
• The temperature control is faulty.
Measure the sample at ambient
temperature within the specified
range (15°C to 35°C). If the
warning nevertheless appears,
contact Eppendorf Service.
Absorbance at the
measuring
wavelength is too
high.
• Turbidity of the measuring solution.
• Optical surfaces of the cuvette are soiled.
• Cuvette has been inserted into the
cuvette shaft facing the wrong direction.
• Too high absorbance of measuring
solution.
Measure again considering the
possible causes.
The calculated
result is negative.
• Measuring solution not prepared
correctly.
• The incorrect factor has been entered
(wrong algebraic sign).
Measure again considering the
possible causes.
At least one of the
results is negative.
• For methods with several results (e.g.,
Dye labels).
• Measuring solution not prepared
correctly.
• The incorrect factor has been entered
(wrong algebraic sign).
Measure again considering the
possible causes.
The result has more
than 6 pre-decimal
places.
• Very high sample concentration.
• Concentration unit does not match the
expected range of the sample
concentrations.
Dilute sample and measure again.
Change the concentration unit
(Parameter
Unit) and measure
again.
The result is more
than 5 % outside of
the standard
concentration
range.
• For methods with evaluation via standard
curves (nonlinear evaluation method):
The sample result is more than 5 %
outside of the standard concentration
range.
Remeasure the sample under
conditions under which the result
is within the range of the standard
concentrations (dilute sample,
modify standard concentrations
and remeasure).
• Calculation not
possible
because of
division by zero.
Absorbance
result is zero.
• Calculation
error. Division
by zero.
• The evaluation required dividing by an
absorbance result with the value of
"zero". This is not mathematically
permissible.
Examples: Calculation of a factor at
one-point calibration; calculation of a
260/280 ratio with nucleic acid
measurements.
Check the reagents and samples
used and repeat the measurement.
Calculation not
possible because of
division by zero.
Absorbance result
or parameter
formula
b is zero.
• An absorbance result was divided by a
"zero" value during the evaluation of a
Division type method (Dual wavelength
method group). This is not
mathematically permissible.
Check the reagents and samples
used and repeat the measurement.
Do not enter "zero" as a value for
the Formula
b parameter.
Problem
Cause
Solution