Maintenance
Eppendorf BioSpectrometer

®

kinetic

English (EN)

66

3. Clean the cover with a cloth or lint-free cotton swab dampened with a mild cleaning agent.

4. Slide the locking pin back into the housing as far as it will go.

The locking pin has completely disappeared in the housing.

8.2

Disinfection/Decontamination

1. Clean the device with a mild cleaning agent before the disinfection (see Cleaning on p. 65).

2. Choose a disinfection method which corresponds to the legal regulations and guidelines in place for

your range of application.

3. For example use alcohol (ethanol, isopropanol) or other alcoholic disinfectants.

4. Wipe the surfaces with a cloth which you have moisturized with a disinfectant.

5. If the cuvette shaft cover needs to be removed for the disinfection, proceed as follows for the

disassembly and assembly (see Cleaning the cuvette shaft cover on p. 65).

6. You can use spray disinfection to disinfect the disassembled cuvette shaft cover.

8.3

Checking the device

Requirements

• Observe the ambient conditions (see Ambient conditions on p. 85).
• Carry out an inspection at approx. 20 °C. Avoid temperature variations (e.g, due to opened windows).
• The filter may only be removed from the filter box for a short period of time, and the filter surfaces must

be protected from contamination or damage.

• Protect the filter from dust, heat, liquids and aggressive vapors.
• The filter is inserted so that the label, with the name of the filter, points toward the detector.

When the spectrometer unit is checked: label points forward.

• The cuvette shaft is free from contamination.

When the photometer is not being used, close the cuvette shaft using the blue cuvette shaft
cover to protect it from dust and other contamination.

DANGER! Electric shock as a result of penetration of liquid.

 Switch off the device and disconnect the power plug before starting cleaning or

disinfection work.

 Do not allow any liquids to penetrate the inside of the housing.
 Do not spray clean/spray disinfect the housing.
 Only plug the device back in if it is completely dry, both inside and outside.