Troubleshooting

Eppendorf BioSpectrometer

®

kinetic

English (EN)

The coefficient of
determination for
the regression
evaluation of the
standard series is
< 0.8.

• For methods with evaluation of standard

series via the regression procedure: If the
regression evaluation for the standard
series was nonlinear, but the standard
evaluation was accepted by the user, a
warning appears after samples have been
measured.

 Use the sample results with the

reservation mentioned or repeat
the measurement of the standard
series and samples.

Scan: Some of the
measured
absorbances are
too high and are
not displayed.

• For at least one scan wavelength, the

absorbance exceeded the measuring
range.

• Turbidity of the measuring solution.
• Measurements at the limits of the

photometric measuring range.

 If the non-displayed areas of the

scan are relevant: Dilute the
sample or remove the dilution via
centrifugation and repeat the
measurement.

The measurement
is not complete.

• Kinetic procedure: You have prematurely

canceled the measurement using the
[Stop] softkey. If at least 2 measuring
points are available, a result is calculated
and displayed.

 Accept the measuring result with

the reduced measuring time or
measure again with a longer
measuring time.

The kinetics are
nonlinear: The
coefficient of
determination is
< 0.95.

• Kinetic procedure with "linear

regression" measuring procedure: The
coefficient of determination for the
regression evaluation indicates a
significant deviation of the measuring
points from the regression line.

• Turbidity of the measuring solution.
• Measurements at the limits of the

photometric measuring range.

• Activity concentration of the enzyme too

high.

 Accept measuring result or

remeasure the sample. Before
repeating the measurement,
evaluate the reason for the
non-linearity and act accordingly
(e.g., clear the measuring solution
by centrifugation or dilute the
sample).

The kinetics are
nonlinear for at
least one standard:
The coefficient of
determination is
<0.95.

• Kinetic procedure with "linear

regression" measuring procedure and
evaluation procedure via standards: The
coefficient of determination for the
regression evaluation of at least one
standard measurement indicates a
significant deviation of the measuring
points from the regression line.

• Turbidity of the measuring solution.
• Measurements at the limits of the

photometric measuring range.

• Activity concentration of the enzyme too

high.

 Accept measuring result or

remeasure the standard. Before
repeating the measurement,
evaluate the reason for the
non-linearity and act accordingly
(e.g., clear the measuring solution
via centrifugation or use the
standard with low concentration).

Problem

Cause

Solution