This chapter provides the technical data of the HMT310.

Specifications

Table 19

Relative Humidity Specifications

Property

Description / Value

Measuring range

0 ... 100 %RH

Sensor

Vaisala HUMICAP

®

180R

Vaisala HUMICAP

®

180RC

Vaisala HUMICAP

®

180VHP


Vaisala HUMICAP

®

180VHPC

for typical applications

for applications with chemical purge

and/or warmed probe

catalytic sensor for environments with

vaporized hydrogen peroxide (H

2

O

2

)

catalytic sensor with chemical purge

for H

2

O

2

environments

Accuracy (including nonlinearity,

hysteresis and repeatability)

at +15 ... +25 C

at -20 ... +40 C

at -40 ... +180 C

± 1 % RH (0 ... 90 %RH)

± 1.7 % RH (90 ... 100 %RH)

± (1.0 + 0.008 × reading) %RH

± (1.5 + 0.015 × reading) %RH

Factory calibration uncertainty

(+20 °C)

±0.6 % RH (0 ... 40 %RH)

±1.0 % RH (40 ... 97 %RH)

(Defined as ± 2 standard deviation

limits. Small variations possible, see

also calibration certificate.)

Response time (90%) for

HUMICAP

®

180R and

HUMICAP

®

180RC at +20°C in

0.1 m/s air flow

17 s with grid filter

50 s with grid and steel, netting filter

60 s with sintered filter