Chapter 6 ___________________________________________________ Calibration and Adjustment

VAISALA ________________________________________________________________________ 69

CHAPTER 6

CALIBRATION AND ADJUSTMENT

This chapter describes the relative humidity and temperature adjustment

procedures. Note that after adjustment, the original calibration certificate

shipped with the product is not valid anymore.

Vaisala recommends using the calibration services offered by Vaisala.

Vaisala Service Centers perform calibrations and adjustments as well as

repair and spare part services, and offer accredited calibrations and

maintenance contracts.

Calibration services can also be ordered online at

http://store.vaisala.com

.

CAUTION

If you think the HMT310 is not measuring correctly, calibration and

adjustment is not the first thing to do. Try the following first:
- Make sure nothing is interfering with the measurement: heat sources.

temperature differences, or condensation

- Check that there is no moisture on the probe. If the sensor has become

wet, you must allow it to dry before you can measure.

- Always wait for the measurement to stabilize.